Error Messages
Possible cause Check Remedy
KMC1
Message Meaning
Run electronics diagnostics
Output voltage V3
defective
LED -22-LD37 not lit Check fuse -22-F85
Power supply voltage
ce
ntral electrical system defective
LED -22-LD60 not lit Check fuse -22-F60
LED -22-LD63 not lit Check fuse -22-F63
Check fuse -22-F92
Fuse F12 in KMC1 defective KMC1 LD40 not lit Check fuse F12 in KMC1
Wiring of release switch defective Check wiring
1105 KMC1 voltage V3
defective!
Error: KMC1
Output voltage V3
Release switch defective Replace release switch
GAL component release logic wrong
(wrong GAL component)
Check release logic, and
replace GAL component, if necessary
GAL component defective Replace GAL component
Charge indicator lamp defective Check charge indicator lamp,
Check wiring
Dynamo defective Check the excitation voltage, check the
wiring, replace the dynamo
Battery dead Charge battery, check battery acid,
replace battery
Internal error - KMC1 Replace KMC1