Error Messages
Check Remedy
DIOM
Message Meaning Possible cause
Run electronics diagnostics
Short circuit/broken cable in the wiring
for the valve
Check wiring to valve
Valve plug defective Replace valve plug
Coil for solenoid valve defective Replace the solenoid valve
Internal error DIOM Replace DIOM
Run electronics diagnostics
Short circuit/broken cable in the wiring
for the valve
Check wiring to valve
16085 DIOM DIG12
status
Error: Status
16084 DIOM DIG11
status
Error: Status
for the valve
Valve plug defective Replace valve plug
Coil for solenoid valve defective Replace the solenoid valve
Internal error DIOM Replace DIOM