Error Messages
Check Remedy
DIOM
Message Meaning Possible cause
Run electronics diagnostics
Short circuit/broken cable in the wiring
for the valve
Check wiring to valve
Valve plug defective Replace valve plug
Coil for solenoid valve defective Replace the solenoid valve
Internal error DIOM Replace DIOM
Run electronics diagnostics
Short circuit/broken cable in the wiring
for the valve
Check wiring to valve
16053 DIOM DIG6 status Error: Status
16052 DIOM DIG5 status Error: Status
for the valve
Valve plug defective Replace valve plug
Coil for solenoid valve defective Replace the solenoid valve
Internal error DIOM Replace DIOM
Run electronics diagnostics
Short circuit/broken cable in the wiring
for the valve
Check wiring to valve
Valve plug defective Replace valve plug
Coil for solenoid valve defective Replace the solenoid valve
Internal error DIOM Replace DIOM
16054 DIOM DIG7 status Error: Status