BIO RAD 400-0174 Instrucciones de operaciones Pagina 449

  • Descarga
  • Añadir a mis manuales
  • Imprimir
  • Pagina
    / 485
  • Tabla de contenidos
  • MARCADORES
  • Valorado. / 5. Basado en revisión del cliente
Vista de pagina 448
Error Messages
Check Remedy
DIOM
Message Meaning Possible cause
Short circuit/broken cable in the wiring
for the valve
Check wiring to valve
Valve plug defective Replace valve plug
Coil for solenoid valve defective Replace the solenoid valve
Internal error DIOM Replace DIOM
Run electronics diagnostics
Short circuit/broken cable in the wiring
for the valve
Check wiring to valve
Valve plug defective
Replace valve plug
16082 DIOM DIG9 status Error: Status
16081 DIOM PWM12
status
Error: Status
Valve plug defective
Replace valve plug
Coil for solenoid valve defective Replace the solenoid valve
Internal error DIOM Replace DIOM
Run electronics diagnostics
Short circuit/broken cable in the wiring
for the valve
Check wiring to valve
Valve plug defective Replace valve plug
Coil for solenoid valve defective Replace the solenoid valve
Internal error DIOM Replace DIOM
16083 DIOM DIG10
status
Error: Status
Vista de pagina 448
1 2 ... 444 445 446 447 448 449 450 451 452 453 454 ... 484 485

Comentarios a estos manuales

Sin comentarios