Error Messages
Check Remedy
DIOM
Message Meaning Possible cause
Run electronics diagnostics
Short circuit/broken cable in the wiring
for the valve
Check wiring to valve
Valve plug defective Replace valve plug
Coil for solenoid valve defective Replace coil
Solenoid valve defective Replace the solenoid valve
Internal error DIOM Replace DIOM
Run electronics diagnostics
16040 DIOM PWM5
current
Error: Over-current
DIOM PWM6
16041
Error: Over-current
Run electronics diagnostics
Short circuit/broken cable in the wiring
for the valve
Check wiring to valve
Valve plug defective Replace valve plug
Coil for solenoid valve defective Replace coil
Solenoid valve defective Replace the solenoid valve
Internal error DIOM Replace DIOM
DIOM PWM6
current
16041
Error: Over-current