Error Messages
Run electronics diagnostics
Power supply voltage
KMC1 defective
LED -22-LD61 not lit Check fuse -22-F61
Power supply voltage
ce
ntral electrical system defective
LED -22-LD60 not lit Check fuse -22-F60
Check fuse -22-F92
KMC1 LD36 not lit
Wiring defective Check wiring
Short circuit in the wiring to a 12V
sensor
Check wiring
Possible cause Check Remedy
KMC1
Message Meaning
1100 Error:
Electronics voltage –
voltage out of range
KMC1 fault
voltage electronics
sensor
Charge indicator lamp defective Check charge indicator lamp,
Check wiring
The controller of the dynamo is
defective
Check the excitation voltage;
Check wiring
Dynamo defective Replace the dynamo
Battery dead Charge battery, check battery acid,
replace battery
Battery change-over relay defective Replace battery change-over relay
Internal error - KMC1 Replace KMC1