4 | Installing ChromLab Software
94 | NGC Chromatography Systems and ChromLab Software
3. Select Synchronize with detector.
4. Refer to the following table to specify the inner diameter of the tubing in use.
Tubing Color Inner Diameter
Orange 0.02" (0.5 mm)
Green 0.03" (.75 mm)
Clear 0.062" (1.6 mm)
5. If your system includes the BioFrac fraction collector, select the checkbox
Include BioFrac - 12
μl and specify the tubing length between the UV outlet port
and the fraction collector drophead.
6. If your system includes an outlet valve, select the checkbox Include Outlet
valve - 50
μl and specify the tubing length between the UV outlet port and
outlet valve.
7. If your system includes both the BioFrac fraction collector and an outlet valve,
select both checkboxes. In the Include Outlet valve - 50
μl section, specify the