5 | Plumbing and Priming the NGC Chromatography Systems
122 | NGC Chromatography Systems and ChromLab Software
b. (Optional) Do one of the following:
Connect Tube #4 from port 1 to the Common inlet port on the BioFrac
fraction collector diverter valve.
Connect Waste tubing to port 1 and immerse in a waste container.
c. Connect outlet tubing to the remaining ports as needed.
2. If your system has two outlet valves:
a. Connect tubing from the Out port on the conductivity monitor or pH
monitor to the In port on outlet valve 1 (O1).
b. (Optional) Do one of the following:
Connect Tube #4 from port 1 on O1 to the Common inlet port on the
BioFrac fraction collector diverter valve.
Connect Waste tubing to port 1 on O1 and immerse in a waste
container.
c. Connect tubing from port 12 on O1 to the In port on outlet valve 2 (O2).
d. Connect outlet tubing to all remaining outlet ports as needed.
3. Immerse the outlet tubing in fraction collection containers.