5 | Plumbing and Priming the NGC Chromatography Systems
146 | NGC Chromatography Systems and ChromLab Software
Calibrating the Conductivity Monitor
Flow conductivity standard from any commercial supplier through the conductivity
flow cell until a stable reading is obtained.
To calibrate the conductivity monitor
1. On the NGC touch screen, select Calibrate on the ChromLab dropdown menu.
2. Select Conductivity Monitor on the dropdown list.
3. Follow the steps in the Conductivity Monitor dialog box.
4. Verify that Status (at the bottom of the dialog box) is Calibrated.