Bio-rad ChromLab™ Software Manual de usuario Pagina 171

  • Descarga
  • Añadir a mis manuales
  • Imprimir
  • Pagina
    / 178
  • Tabla de contenidos
  • SOLUCIÓN DE PROBLEMAS
  • MARCADORES
  • Valorado. / 5. Basado en revisión del cliente
Vista de pagina 170
Using the C-96 Autosampler with NGC Systems
Installation Guide | 169
To prime and purge the C-96 autosampler
Note: Before priming the autosampler, ensure that any column is offline.
1. Immerse the inlet tubing in a container of buffer or HPLC grade (filtered,
degassed) or other high quality water
.
2. Use a syringe to prime the pumps. See the section Priming and Purging the
System on page 123 for specific information.
3. On the ChromLab™ computer or the touch screen, start ChromLab in Manual
mode.
4. Set the fluidic scheme to display the autosampler and sample inject valve.
5. In the fluidic scheme, touch or click the Sample Inject Valve module to open its
dialog box.
Autosampler
Sample Inject
valve
Vista de pagina 170

Comentarios a estos manuales

Sin comentarios