Priming and Purging the System
Installation Guide | 137
5. Insert a 30 ml syringe into the sample pump’s priming port.
6. Slowly withdraw the
plunger to withdraw liquid f
rom the port
until
air no
longer enters the syringe (approximately 10 ml).
7. Remove the syringe from the priming port and discard the liquid into waste.
8. Repeat steps 4–7 for the remaining sample inlet ports.