3 | System Control
38 | NGC Chromatography Systems and ChromLab Software
Changing Chromatogram View Settings
You can change trace display attributes and other run view settings in the System
Control window. Changes you make to the following settings are saved and used in
the display of subsequent runs:
Trace Show/Hide state
Trace color
Trace y-scale range (Min/Max y values)
Active x-scale units
Show/Hide state of events, fractions, errors, and annotations
Delta Col
Pressure
Column Switching Valve Available when at least one column switching
valve is present
Tip: When two or mor
e column switching
valves are present, the traces appear in
different shades of blue on the
chromatogram. The trace number
corresponds to the valve number in the fluidic
scheme. For example,
Col Pressure 1 in the
legend and the chromatogram corresponds
to C1 in the fluidic scheme.
Sample Pump
Pr
essure
Sample Pump Sample pump pressure (psi or MPa)
Temperature Single-Wavelength UV
and Multi-W
avelength
UV/Vis Detectors
The temperature of the fluid, read at the Single-
Wavelength UV or Multi-Wavelength UV/Vis
module
Flow Rate System Pump Current system flow rate (ml/min)
Sample Pump
Flow Rate
Sample Pump Sample pump flow rate (ml/min)
Table 1. Trace Definitions, continued
Trace Module Explanation