7 | Evaluating Results
228 | NGC Chromatography Systems and ChromLab Software
Automatic Peak Integration Settings
In the Peak Integration pane, the Auto Integration tab displays settings that affect
baseline calculation and peak detection. Changing these settings enables you to
optimize the way peaks are detected and recognized, specific to your data.
Note: By default most tab sections are collapsed. Expanding the sections
displays the options shown in the following illustration.
Traces
A list of shown UV traces grouped by run. By default all
shown UV traces are selected for peak integration.
You can exclude traces from integration by clearing their
checkboxes.
Baseline Parameters
You can select from two baseline calculation methods.
By Best Fit — the baseline is calculated from the data.
The baseline curve connects the low points of the data.
This setting describes the radius of a disk r
olling along
the curve from underneath.
By default, the best fit radius parameter is set to 8. This
value can to be set fr
om 1 to 10. The smaller the radius,
the mor
e low points on the curve it touches.
By Offset — the baseline is a straight horizontal line
that has a constant of
fset (y) value. The default value is
zer
o, but you can specify a different offset value.