Valves
Instrument Guide | 4 1
Sample Pump Direct Inject/System Pump Waste
Directs system pump flow to waste
(W2)
Directs sample pump flow directly
onto the column to load large
volumes of sample
Chromatogram x-axis based on
sample pump flow rate
Sample Pump Load Loop/System Pump to Column
Directs system pump flow directly
onto the column
Directs sample pump flow into the
loop to load the loop with sample.
Excess sample flows out through
waste port (W1)
Chromatogram x-axis based on
system pump flow rate
Valve Position Explanation