Bio-rad ChromLab™ Software Manual de usuario Pagina 239

  • Descarga
  • Añadir a mis manuales
  • Imprimir
  • Pagina
    / 306
  • Tabla de contenidos
  • MARCADORES
  • Valorado. / 5. Basado en revisión del cliente
Vista de pagina 238
Other Components
Instrument Guide | 237
To remove the backpressure regulator
1. Disconnect the tubing from the top and bottom of the installed backpressure
regulator.
2. Store the backpressure regulator in a safe place.
3. Do one of the following:
Attach a Tefzel union to the open tubing in place of the backpressure
regulator.
Replace the two open tubes with a single tubing line.
Attach one end of the tubing to the top of the conductivity monitor and the
other end to either the pH flow cell or the BioFrac™ fraction collector.
Vista de pagina 238
1 2 ... 234 235 236 237 238 239 240 241 242 243 244 ... 305 306

Comentarios a estos manuales

Sin comentarios