Bio-rad ChromLab™ Software Manual de usuario Pagina 215

  • Descarga
  • Añadir a mis manuales
  • Imprimir
  • Pagina
    / 306
  • Tabla de contenidos
  • MARCADORES
  • Valorado. / 5. Basado en revisión del cliente
Vista de pagina 214
The Detectors
Instrument Guide | 213
11. Insert the cable plug and tighten the collar.
Tip: The deuterium lamp plug has seven connector pins, the tungsten
lamp has five.
12. Replace the access door and tighten the screw with a Phillips screwdriver.
13. Slide the multi-wavelength UV/Vis module into its bay and fasten it in place by
tightening the two screws.
14. Reconnect the tubing lines to the top and bottom of the UV cell.
15. Reconnect the tubing lines to the conductivity monitor.
16. Restart the NGC instrument.
Note: The multi-wavelength UV/Vis detector will go through an automated
calibration routine when the system is restarted.
Vista de pagina 214
1 2 ... 210 211 212 213 214 215 216 217 218 219 220 ... 305 306

Comentarios a estos manuales

Sin comentarios