Bio-rad ChromLab™ Software Manual de usuario Pagina 135

  • Descarga
  • Añadir a mis manuales
  • Imprimir
  • Pagina
    / 306
  • Tabla de contenidos
  • MARCADORES
  • Valorado. / 5. Basado en revisión del cliente
Vista de pagina 134
Priming and Purging the Systems
Instrument Guide | 133
Purging the NGC Discover System
To purge the buffer blending valve
1. Open the System Pump dialog box and select Priming on the Mode dropdown
list.
Note: When in Priming mode, the sample inject valve switches the system
pump flow to waste (W2). After the process is completed, the sample inject
valve switches the system pump to manually load the column through the
loop.
To purge an individual line of air bubbles, proceed to step 2 on page 134.
To purge all lines of air bubbles sequentially, proceed to step 3 on
page 134.
Vista de pagina 134
1 2 ... 130 131 132 133 134 135 136 137 138 139 140 ... 305 306

Comentarios a estos manuales

Sin comentarios