6. Run the same protocol on partition “n” of 4 for all 8 spots. Record the
partition number on the usage chart provided with the product packaging.
7. Make sure that a peak intensity of 50–500 µA is visible on each spot. If peak intensities
appear out of this range, change the laser energy and run the test again.
16