Service and Certification
16 | S3 Biosafety System
S3 Biosafety System Class I Certification Form
S3 Biosafety System Class I Certification Form
Customer Name Phone
Company
Location
Enclosure SN
Inflow Velocity
Airflow readings, FPM __________________________ __________________________
__________________________ __________________________
__________________________ __________________________
__________________________ __________________________
__________________________ __________________________
Calculated average airflow _________________________ FPM
Average airflow >135 FPM? ______ PASS or ______ FAIL
Measured area ______ ft
2
Area of perimeter base slots ______ 0.15 _________ ft
2
Total area (measured area + perimeter slots) ______ ft
2
Exhaust volume (average airflow x total area) _________________________ CFM
Leak Integrity
Upstream concentration ______________________________
Upstream concentration >10 µg/L? ______ PASS or ______ FAIL
Exhaust discharge concentration ______________________________
Exhaust discharge <0.01% upstream concentration? ______ PASS or ______ FAIL
Smoke Containment
Smoke drawn into enclosure? ______ PASS or ______ FAIL
Certifier (Printed Name)
Certifier (Signature)
Date of Certification
Comentarios a estos manuales