Instruction Manual | 83
9
Automatic Startup
The ProSort
can schedule an automatic startup of the system. Upon automatic start up, the
system will verify the stream is coarsely aligned and no clogs are present. If these conditions are
not met, the stream will be shut off automatically to avoid spraying or flooding. A warning will
appear and the system may be manually started to resolve the issue.
9.1 Scheduling an Automatic Startup
There are three ways to schedule an automatic startup.
1. Using the Startup button (provided no scheduled startup has been set). Please refer to
Section 4.3, Daily Startup.
2. During the shutdown procedure. Please refer to Section 8.1, Daily Shutdown.
3. During the decontamination procedure. Please refer to Section 10.7, Decontamination.
Fig. 94. Startup settings.