Bio-rad S3™ Cell Sorter Manual de usuario Pagina 117

  • Descarga
  • Añadir a mis manuales
  • Imprimir
  • Pagina
    / 132
  • Tabla de contenidos
  • SOLUCIÓN DE PROBLEMAS
  • MARCADORES
  • Valorado. / 5. Basado en revisión del cliente
Vista de pagina 116
10.7 Decontamination
Instruction Manual | 107
8. Replace the DI water and sheath fluid containers with at least 1 L of decontamination
solution (Figure 130).
9. Empty waste container and replace.
10. Click the checkmark button to proceed.
The system will drain the internal sheath reservoirs into the waste container (Figure 131).
The system will directly prime the fluidics with the decontamination fluid (Figure 132).
Fig. 131. Decontamination window
while draining internal system.
Fig. 132. Decontamination window while
priming system.
Fig. 133. Decontamination window while priming rinse filter and nozzle path.
Fig. 130. Decontamination instructions to prepare the system.
Vista de pagina 116
1 2 ... 112 113 114 115 116 117 118 119 120 121 122 ... 131 132

Comentarios a estos manuales

Sin comentarios