Appendix A
Flow Cell Replacement
1. Remove the retaining clip from the bottom of the optics module and
pull the luer connectors away from their holders as shown in Figure
A-1.
2. Unscrew thumbscrew and remove the old flow cell.
3. Reinsert the new flow cell following the above steps in reverse
order. (Do not over-tighten the thumb screw.)
Warning: To avoid exposure to UV radiation, make sure that the
UV lamp is off when changing filters and flow cells.
Fig. A-1. Flow cell replacement.
!
Retaining Clip
Thumbscrew
Flow Cell
Luer Connectors
M7318160G.qxd 05/08/2003 2:05 PM Page 14