Appendix A
Fuse Replacement
Figure A. Fuse removal.
1. Disconnect unit from power source.
2. Remove the fuse cover with a small-blade screwdriver or similar tool.
3. Pull the fuse holder out of the fuse compartment and, if necessary,
replace the fuse. Use 5 mm x 20 mm fuse, type T, 1A.
4. Reinsert the fuse holder, with fuse, into the fuse compartment.
Appendix B
Rear Panel Connections
The signal output socket is an 8-pin circular mini-DIN. The fol-
lowing information is provided for those wishing to interface the Model
EG-1 Econo Gradient Monitor with equipment such as integrators or chart
recorders which do not utilize banana plug connectors.
Pin Function
1 No Contact
2 No Contact
3 No Contact
4 No Contact
5 Chart Recorder (+)
6 Chart Recorder (-)
7 No Contact
8 No Contact
Pin configuration