Model 680
Microplate Reader
Instruction Manual
Catalog Numbers 168-1000
168-1001
168-1002
168-1003
Vista de pagina 0
1 2 3 4 5 6 ... 67 68

Indice de contenidos

Pagina 1 - Instruction Manual

Model 680 Microplate Reader Instruction Manual Catalog Numbers 168-1000 168-1001 168-1002 168-1003

Pagina 2 - Warranty Statement

6 Section 3. Instrument Set-up 3.1 Initial Start-up 1 Place the instrument on a clean, sturdy table or bench. It is important to keep the instrum

Pagina 3 - Safety:

7 Section 4. Operation 4.1 Instrument Overview The Model 680 Microplate Reader has built-in software that allows the user to set the plate reading

Pagina 4 - Table of Contents

8 Protocol Security Filters Date Set Lab name Cutoff Report Limit STDs Mode Mapping Kit name Not Use Constant Control FormulaRatio Ranged Cutoff Sing

Pagina 5 - Section 1. Introduction

9 .Func key to switch from [ABC] to [abc] to [sym] key to move cursor .Func key to switch from [ABC] to [abc] to [sym] key to move cursor .Fu

Pagina 6

10 4.2 Functions Overview 4.2.1 Calculations The Microplate Reader uses Beer's Law to calculate the absorbance value of each well. Beer's

Pagina 7

11 in this memory. The first time the instrument is turned on, or after a battery failure, the following default information will be held in the mem

Pagina 8

12 Blank mean = X/n S.D. = [{X^2 – n*(Blank mean)^2}/{n-1}]^1/2 Where: S.D. = Standard deviation X = Sum total of the raw absorbance for each blank X

Pagina 9 - 2.3 Membrane Keypad

13 Upper cutoff absorbance = Positive absorbance + ((Gray zone/100) * Positive absorbance) Lower cutoff absorbance = Positive absorbance – ((Gray zo

Pagina 10 - 3.1 Initial Start-up

14 The result absorbance value of the formula calculation and the gray zone value that is entered by the user are used for the cutoff. The upper and

Pagina 11 - 4.1 Instrument Overview

15 i. 5P Logistic Rodbard ii. 4P Logistic Rodbard iii. 5P Logistic Cook-Wilkenson iv. 4P Logistic Cook-Wilkenson v. 5P Exponential vi. Sigmoid Logist

Pagina 12

i Warranty Statement This warranty may vary outside of the continental United States. Contact your local Bio-Rad office for exact terms of your war

Pagina 13

16 Negative cutoff = (Negative Control average + 0.210) * 0.900 TeSeE eq: Positive cutoff = Negative Control average + 0.210 Negative cutoff = Negat

Pagina 14 - 4.2 Functions Overview

17 4.2.7 Printer The Model 680 Microplate Reader has a built-in printer interface connector (25 pin D-sub) or an optional factory built-in printer. T

Pagina 15 - Abs = Raw – Blank mean

18 4.3 Detailed Operation 4.3.1. Initial Start-up 01:Kinetic-1 M405(1)R605(6)30readShake:999s,Mid,Every38.2/40

Pagina 16

19 4.3.2. Configuring the system Note: To return to the previous screen, use the back/left arrow key.

Pagina 17

20 1:400nm 5:650nm 2:450nm 6:...nm 3:475nm 7:...nm 4:650nm 8:...nm Filter Day/Month/Ye

Pagina 18

21 4.3.3. Printing Reports and Protocol information Print menu » Reports •Protocol PRINT MENU Print Menu This menu

Pagina 19

22 4.3.4. Memory Recall operation 01 20/05/00 14:00 02»*21/05/00 17:00 03 20/05/00 18:00 ↓04»

Pagina 20 - Remote mode

23 4.3.5. Editing Kinetic protocol Kinetic Protocol Menu Select “Protocol” in the Edit Menu and p

Pagina 21

24 Kit name for #01 »G.A.P. TEST IgG« ↑ [ABC] Kit name Calculation Mapping As

Pagina 22 - 4.3.1. Initial Start-up

25 4.3.6. Checkmark value setting (by Administrator only) » Serial #/Report •Linearity ref. •Max

Pagina 23

ii Regulatory Notices Electro Magnetic Compatibility: This instrument is designed to meet the Class-A emissions and immunity levels of the EN61326

Pagina 24

26 4.3.7. Editing End Point Protocol Upper Limit= 2.000 Lower Limit= 0.000 input range: 0–3.500 U

Pagina 25 - External Printer setting

27 4.3.8. Cutoff setting procedure Cutoff setting Menu This menu is entered by choosing "Cutoff" in the End Point Protocol Menu. It is us

Pagina 26

28 b. Cutoff Control setting Selection of Single/Ranged Cutoff Control This menu screen selects Single or Ranged cutoff type. The cutoff type marked

Pagina 27 - Edit Menu

29 c. Cutoff Formula setting Selection of Formula This menu screen selects a formula for cutoff calculation. The formula marked with “[ ]” is the c

Pagina 28

30 d. Cutoff Ratio setting Ratio to Calibrator In this screen, there are three items for setting the ratio to calibrator. The Up and Down arrow key

Pagina 29 - Checkmark Protocol Menu

31 4.3.9. Standard setting procedure Standard curve setting Menu This menu screen is entered by choosing "STDs" in the End Point Protocol

Pagina 30 - Edit Menu

32 b. Curve fit setting Curve fit In this screen, there are two items for setting the curve fit. The Up and Down arrow key moves the cursor to the

Pagina 31 - Ranged Cutoff Constant

33 c. Recall STD curve setting Recall Setting of Standard Curve In this screen, there are two items for setting the Recall STD curve as below. The U

Pagina 32 - Ranged Cutoff Control

34 4.3.10. Mode setting procedure Mode setting Menu This menu screen is entered by choosing the "Mode" in the End Point Protocol Menu sc

Pagina 33

35 d. Incubation Incubation In this screen, the user switches on/off the Incubator and sets the incubation temperature. The temperature setting rang

Pagina 34

iii Table of Contents Section 1. ...Introduction ...

Pagina 35 - Concentration

36 4.3.11. Plate Map setting procedure Select Mapping mode Menu This menu screen is entered by choosing the "Mapping" in the End Point Pr

Pagina 36 - Graph axis

37 Acceptable range of number for each well type: 1. Blank and Empty wells are not acceptable for any number. 2. Sample and Standard wells: 00

Pagina 37 - Storage number selection

38 Warning and Error of Automatic Mapping: Warning screen: If the total number of mapping wells exceeds 96 wells, a warning message in the left figu

Pagina 38 - Read mode

39 4.4 Quick Guide – Reading a plate 1. Turn on the instrument. Allow about 15 seconds for the self-diagnosis. Allow the instrument to warm up fo

Pagina 39 - Incubation

40 4.5 Language choice The Model 680 Microplate Reader has two versions, English and Chinese, in the on-board software. And it allows the user to ch

Pagina 40 - Manual Mapping

41 (5). Retrieval function for plate data The Chinese version does not support this feature. The URPD (Upload Raw Plate Data) and DRPD (Download Raw

Pagina 41 - Automatic mapping

42 Section 5. Instrument service by the user Caution: Electrical shock hazardous! Always disconnect the instrument from the AC powe

Pagina 42

43 Section 6. Troubleshooting and Error Messages The Model 680 Microplate Reader constantly monitors several functions and will display appropriat

Pagina 43

44 7. LCD displays “Filter not installed”. Enter the Edit Menu by pressing Edit key, and set up the filters. 8. Wells have color, but absorbance

Pagina 44 - 4.5 Language choice

45 Section 7. Specifications 7.1 Instrument Specifications Operating panel Membrane pad with 4 arrow keys, 7 function keys and 11 numeral keys Di

Pagina 45

1 Section 1. Introduction The Model 680 Microplate Reader is an eight-channel, vertical pathlength photometer that measures the absorbance of the c

Pagina 46 - 5.2 Changing the Lamp

46 Accuracy ±1.0% or 0.010 from 0 to 3.0 OD at 490nm Linearity ±1.0% from 0 to 2.0 OD, ±2.0% from 0 to 3.0 OD Reproducibility 1.0% or 0.0

Pagina 47

47 7.2 Interface Specifications The following specification describes the syntax and language required for computer control of the reader. 1. SI

Pagina 48

48 7.3 Command Language for Model 680 Microplate Reader 7.3.1 Command Language for Remote control mode 1-1. Syntax SYNTAX: Command: <devic

Pagina 49 - Section 7. Specifications

49 1-2-6 RM (Reset maintenance report) <command>: "RM" Reset maintenance report values. <arguments>: none <response>: &qu

Pagina 50

50 1-2-11. RWELL (Read a plate well) <command>: “RWELL” Read a plate well <arguments>: Single reading:<cl><space><rw>&

Pagina 51 - 7.2 Interface Specifications

51 1-2-14. UPLD (Upload the protocol data) <command>: "UPLD" Upload the protocol data <arguments>: <up load protocol numbe

Pagina 52

52 1-2-20. SSN (Set Serial Number) <command>: "SSN" Set Serial number <arguments>: <Serial number> <Serial number>

Pagina 53

53 1-3. Absorbance Data 1-3-1. Abs data <abs.data>: <header><date time><reading info><Mes. data><cr>[<Ref.

Pagina 54 - 1-2-12. RPLATE (Read plate)

54 1-4. Protocol data All protocol data is ASCII code data. Transmit format: <“,”(0x2c)><protocol data 1>< “,”><protocol dat

Pagina 55

55 (data length max: 4 characters(include decimal point)) 23 Cutoff negative const value “.001 to 9999” (data length max: 4 characters(include decima

Pagina 56

2 1.1 Description and accessories for Model 680 Microplate Reader Catalog Number Description 168-1000 Model 680 microplate reader 168-1001 Model

Pagina 57 - 1-3-1. Abs data

56 [7]: Difference subtraction 0:Even-Odd (column) 1:Odd-Even(column) 2:Even-Odd(row) 3:Odd-Even(row) [8]: Concentration (“ “(0x20) data delimits pri

Pagina 58

57 1-5. “TSE” and “TSE eq” Assay Parameter Format “TSE” and “TSE eq” parameter is same format. All parameter data is ASCII code data. Transmit fo

Pagina 59

58 1-6. Upload Raw Data Format 1-6-1. Header Type Format of the Upload Raw Plate Data <upload raw plate data>: <memory #><data exist

Pagina 60

59 1-6-2-2. Full Data Type Format for Kinetic <upload raw plate data>: <memory #><data exist><date time><plate #><

Pagina 61

60 1-7. Download Raw Plate Data Format All protocol data is ASCII code data. Transmit format: <“,”(0x2c)><plate datum 1><“,”><p

Pagina 62 - 1-6. Upload Raw Data Format

61 Abs. Data Sample of <Mes. data> and <Ref. data> " 0.101 0.102 0.103 0.104 0.105 0.106 0.107 0.108 0.109 0.110 0.111 0.112"&l

Pagina 63

62 (2) Reading number= "1" to "30" 5 Plate reading date <year>”/”<month>“/”<day><space><hour>”:”&l

Pagina 64

63 7.3.2 Error Codes Error code Error Cause Recovery 0000 No error 8071 Invalid Command The received data is not a command for the instrument 8

Pagina 65

64 4000155 Rev F

Pagina 66

3 Section 2. Product Description 2.1 Contents of Shipping Carton The shipping carton contains the following items: • Model 680 Microplate Read

Pagina 67

4 2.2 External Features 1. Release latch for rear cover 4. Rear cover 2. Liquid crystal display (LCD) 5. Printer slot (wh

Pagina 68

5 2.3 Membrane Keypad 1. Main Returns to MAIN SCREEN. 2. Start/Stop Initiates plate reading using current active pr