Bio-rad Microplate Manager Software Manual de usuario Pagina 55

  • Descarga
  • Añadir a mis manuales
  • Imprimir
  • Pagina
    / 112
  • Tabla de contenidos
  • MARCADORES
  • Valorado. / 5. Basado en revisión del cliente
Vista de pagina 54
Analyzing Data Using Cutoffs
47
Reference can be made to standard or sample IDs that have been defined in the
Template setup
Standards: enter ! followed by concentration
Unknowns: enter identifier
Blanks: enter #
In all Cutoff equations, Microplate Manager 6.0 uses the averaged value of that
group.
Examples of valid expressions:
U01
(U01+.1)/2
(!100-!10)*2
Cutoff Examples for Assays with Controls
Positive & Negative Controls – positive (POS) & negative (NEG) control well(s)
defined on template
Low Cutoff: NEG
High Cutoff: POS
% B/Bo – non-specific binding blank (NSB) & maximum or Bo well(s) (MAX)
defined on template
Low Cutoff: NSB
High Cutoff: MAX
% Calibrator –blank (#) & calibrator (CALB) well(s) defined on template
Low Cutoff: #
High Cutoff: CALB
% Control - non-specific binding blank (NSB) & control well(s) (CNTRL) defined
on template
Low Cutoff: NSB
High Cutoff: CNTRL