PowerPac™ HC Power SupplyInstruction ManualCatalog Number164-5052

Pagina 2

Section 3Setup and Operation3.1 Setup and Programming a RunThis section guides the user through setting up the power supply for a run, connecting the

Pagina 3 - List of Figures

4. Enter constant value Press the Values keys to increase ordecrease the selected value. The keyschange the value in increments of 1,however after a

Pagina 4

STEP PROCEDURE DESCRIPTION2. Editing Use the cursor keys and values keys tochange parameters. For multiple-stepruns, future steps may be accessed and

Pagina 5 - Important

Rapid Resistance Change Detection (RRCd)The Rapid Resistance Change Detection safety feature alerts the user when theresistance load detected by the P

Pagina 6

Section 4Maintenance and Troubleshooting4.1 MaintenanceThe PowerPac HC requires little maintenance to assure reliable operation. Toclean the case, fi

Pagina 7 - Introduction

If a power failure occurs during a run, there are several possible scenarios whichare described as follows:Message Displayed ExplanationRUN ABORTED PF

Pagina 8 - 1.2 Unpacking

Problem Cause SolutionE09 RESISTANCE Change in load Check and correct any potentialCHANGE DETECTED resistance detected. resistance problem. The PowerP

Pagina 9 - Control Features

4.2.4 Bio-Rad Serviceable Error MessagesError code Cause SolutionERROR 03 Output regulation Restart the unit by turning the power* POWER OFF/ON pro

Pagina 10 - Setup and Operation

Appendix ASpecificationsOutput specifications 250 V, 3.0 A, 300 WOutput range (programmable) 5–250 V, fully adjustable in 1 V increments0.01–3.0 A, fu

Pagina 11

Appendix BWarranty and Ordering InformationWarrantyThe PowerPac HC power supply is covered by a standard Bio-Rad Laboratorieswarranty. Contact your lo

Pagina 17 - 4.3 Replacing a fuse

Life ScienceGroup Sig 12124006222 Rev D US/EGBio-Rad Laboratories, Inc.Web site www.bio-rad.com USA 800 424 6723 Australia 61 2 9914 2800 Au

Pagina 18 - Specifications

Table of ContentsPageSafety ...iSection 1 Introduction ...

Pagina 20

SafetyCaution/WarningPowerPac power supplies use high output voltages that are electrically isolatedfrom earth ground through a protective impedance t

Pagina 22

Section 1Introduction1.1 OverviewThe PowerPac HC power supply is designed to provide constant voltage, current orpower for electrophoresis applicatio

Pagina 23

2The PowerPac HC power supply has the following features:• Programmable constant voltage, constant current or constant power operationwith automatic c

Pagina 24 - Laboratories, Inc

Section 2Control FeaturesFig. 4. Front Panel.Key Description Run/Pause Key• Starts or pauses a run. Pausing allows editing of the run parametervalue